13. AC TRUE RMS VOLTS & AMPS INPUT
AC voltage or current measurement utilizes the True 
RMS signal conditioner board which uses precision 
circuitry to compute the root-mean-square of complex 
waveforms from 10 Hz to 10 kHz. Accurate measure-
ments are obtained with spikes up to 3 times the 
maximum of each range. The input can be AC 
coupled to read only the AC component, such as 
ripple on a power supply, or DC coupled to read AC 
plus DC. The board needs to be configured via 
jumpers for the desired voltage or current range, and 
for AC or DC coupling. All signal ranges are factory 
calibrated with calibration factors stored in EEPROM. 
The meter software recognizes the board and will bring up the appropriate menu items for it; 
however, it does not recognize the jumper settings. These need to be set manually. Please 
see further manual sections for setup of the following features: relay output (16), analog output 
(17), and communications (18). 
 
 
 
 RANGE SELECTION VIA JUMPERS 
1.
Use 2.5 mm (0.1") jumpers.
2.
Store spare jumpers on unused jumper post.
METER SCALING
Refer to the above tables for the full scale counts (or displayed digits) produced by the 
available full scale input ranges with a scale factor of 1 and an offset of 0. The decimal point 
can be set for direct readout in (milli)volts or (milli)amperes. Decimal point selection does not 
affect the counts. For example, a 20V input may be displayed as 20.000V or 20000 mV. 
Voltage Ranges
Full Scale Input
Counts
Jumpers
200 mV 
2V 
20V 
200V 
300V (UL) 
600V (not UL) 
20000 
20000 
20000 
20000 
3000 
6000 
j
c, g, h
c, i
c, k
c, m 
c, m 
Current Ranges
Full Scale Input
Counts
Jumpers
2 mA 
20 mA 
200 mA 
5A 
20000 
20000 
20000 
5000
l, k
b, m 
a, m 
c, d, e, m
AC or DC Coupling
Coupling Type
Jumpers
DC coupling for AC + DC 
AC coupling for AC only 
f
none