P7.0: Dissolved Oxygen Parameters (DPr)
This Set Up menu allows you to set the barometric
pressure and the salinity value of the sample to be
measured. You are given the option to use mmHg (Hg)
or kilo Pascal (PA) barometric pressure units. The
Salinity (SAL) value to be entered will be based on ppt
and is available in mg/L (ppm) Set Up menu.
8.8.1
Pressure Setting Adjustment
From the mg/L or ppm measurement mode,
1. Press the SETUP to enter the mg/L or ppm Set
Up mode.
2. Press the ▲ and ▼ keys to scroll through the Set
Up menu till the display shows “dPr”.
3. Press the HOLD/ENTER key and the meter will
enter the barometric pressure unit selection menu
with the display showing either the “HG” mode or
“PA” mode.
4. Use the ▲ and ▼ keys to toggle between the
barometric pressure units and press the
HOLD/ENTER key to confirm selection.
5. Use the ▲ and ▼ keys to set the pressure value
and press the HOLD/ENTER to confirm the
setting. The meter will show the next display of
Salinity setting “SAL”.
NOTE: If the Set Up menu is derived from other than
the mg/L (ppm) mode, the meter will return to the Set
Up subgroup menu “dPr”.
6. Press CAL key to return to measurement mode,
or continue to make a salinity setting adjustment.
8.8.2
Salinity Setting Adjustment (available in
mg/L or ppm Set Up mode only)
1. Repeat steps 1 to 5 as in the steps of Pressure
Setting Adjustment.
2. Use the ▲ and ▼ keys to enter the salinity of your
solution in ppt (parts per thousand).
3. Press HOLD/ENTER key to confirm the value and
return to the subgroup menu “dPr”.
4. Press CAL key to return to the measurement mode.