Thermo Fisher Scientific CyberScan DO 110 User Manual
Page 23

Instruction Manual
CyberScan DO 110
18
7.3
Print Data from Stored Memory
The DO 110 meter can print data that is stored in the memory. You can print out all
stored data or selected data based on its memory location.
To Print All Stored Data
1. From measurement mode, press PRINT key.
2. Use MI/▲ or MR/▼ key, select “dAt – StO” option and press HOLD/ENTER
key. See Figure 9.
3. At the “SEC – 0” option, use MI/▲ or MR/▼ key to select number of seconds
to print each set of stored data. This setting gives flexibility to match the meter
to a printer’s or computer’s ability (or buffer size) to accept data and prevent
overflow.
4. Press HOLD/ENTER key. The “Prn” primary display will flash until all data is
printed.
5. To stop printing, press PRINT key and exit to measurement mode. See Figure
12.
NOTE: All stored memory will remain in the meter after printing.
0
10
sec
sec
ENTER
HOLD
ENTER
HOLD
MI
MR
Flash until all stored
data are downloaded.
Press PRINT key to exit
to measurement mode
Set timed interval to
download stored data
to printer or computer
MEAS
ATC
°C
mg/l
8.10
24.5
READY
From measurement mode
Figure 12: Print all stored data