P 6.0 – Thermo Fisher Scientific CyberScan DO 110 User Manual

Page 36

background image

Instruction Manual

CyberScan DO 110

31

8.7

P 6.0: Memory Clear (CLr)

Use this parameter to clear all memory values when you need to store a new series
of values. This lets you avoid confusing the old values with the new ones.

NO is the default setting.

NOTE: Selecting YES will wipe out all memory.

From measurement mode:

1. Press the SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view the parameter P6.0.

3. Press the HOLD/ENTER key.

4. Press the MI/▲ or MR/▼ keys to toggle between

NO and YES.

• NO retains current memory

• YES clears all memory.

5. Press the HOLD/ENTER key to confirm selection

and return to the subgroup menu.

6. Press CAL/MEAS key to return to measurement

mode.

ENTER

HOLD

SETUP

MEM

p 6.0

SETUP

MEM

SETUP

MEM

yes

MI

MR

Figure 24: P6.0 - Clear

memory