beautypg.com

Thermo Fisher Scientific CyberScan DO 110 User Manual

Page 22

background image

Instruction Manual

CyberScan DO 110

17

7.2.1

Print Current Data Manually

To manually print any data that is currently being measured:

1. From measurement mode, press HOLD key to freeze

reading. See Figure 10.

2. Print PRINT (

) key.

3. The “Prn – CUr” annunciator will appear and flash for

about 0.5 seconds. The data is transmitted to the printer
or computer.

4. The meter returns to measurement mode.

5. Repeat steps 1 to 3 for next measurement.

7.2.2

Print Data on Timed Interval

This feature is useful to automate printing and downloading ‘live’ measurement
values into computer. The time interval setting can range from
5 to 30 seconds.

1. From the measurement mode, press PRINT key.

2. Select “Dat/Cur” screen and press HOLD/ENTER key.

3. The “Sec 5” screen appears. The default value is 5

seconds time interval.

4. Press MI/▲ or MR/▼ key to set a value suitable for your

application.

5. Press HOLD/ENTER key.

6. The meter goes into measurement mode with MEM

annunciator flashing.

7. At a pre-set time interval, the screen flashes “Prn/CUr”

momentarily (printing data) and switches back to
measurement mode with MEM annunciator flashing.

8. To stop printing, press PRINT key. The MEM

annunciator disappears and the meter is on measurement mode.

Figure 10: Print current data

manually

Flash for 0.5 seconds

MEAS

ATC

°C

mg/l

8.10

24.5

READY

HOLD

MEAS

ATC

°C

mg/l

8.10

24.5

READY

5

10

sec

sec

ENTER

HOLD

MI

MR

Figure 11: Print data on timed

interval