Instruction Manual

CyberScan DO 110

29

8.6

P5.0 : Reset to Factory Default (rSt)

This mode lets you reset all parameters to factory default settings. There are two
levels of reset:

• Calibration Reset: Reset only the calibration

values of % Saturation and mg/L (or ppm).
This clears all previous calibrated values
and reset to factory default.

• User Reset: Clears all data including

calibration, memory and other customised
setup functions and reset it to factory
default.

8.6.1

Calibration Reset

From measurement mode:

1. Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P5.0.

3. Press the HOLD/ENTER key.

4. Press the MI/▲ or MR/▼ keys to toggle

between NO and YES.

• ‘NO’ retains current settings

• ‘YES’ erase calibration settings.

5. Press the HOLD/ENTER key to confirm

selection.

• If ‘NO’ is selected, it will retains current

settings and proceed to next program
P5.2 – User Reset.

• If ‘YES’ is selected, all LCD segments will

lit up for one second, erase calibration
values, and the meter switches to the
measurement mode.

6. Otherwise press CAL/MEAS key to return to

measurement mode without resetting to factory
default.

yes

SETUP

SETUP

SETUP

SETUP

P 5.0

P 5.1

CAL

CAL

CAL

ENTER

HOLD

ENTER

HOLD

MI

MR

Select either "YES" or "nO",

press ENTER key to confirm.

Figure 22: P5.1 - Calibration

reset