Instruction Manual

CyberScan DO 110

21

8.1

Advanced SETUP Mode Overview

Press the SETUP key to enter Setup mode. Press the MI/▲ and MR/▼ keys to scroll
through the sub groups.

SETUP

P 1.0

COF

P1.0: Unit Configuration

P1.1

Ready indicator (On, Off or Auto Hold)

P1.2

Select °C or °F temperature units

P1.3

Select Automatic Temperature Compensation (ATC)

P1.4

Select mg/L or ppm mode (from mg/L or ppm mode
only)

SETUP

pH

CAL

P 2.0

P2.0: View Previous pH Calibration Data (View only)

P2.1

% Saturation calibration point ( from % Saturation
setup) or mg/L(ppm) calibration point (from mg/L or
ppm setup only)


SETUP

ELE

P 3.0

P3.0: View Electrode Data (View only)

In % Saturation mode:
EL.1 Electrode

Slope

Factor

EL.2

% Saturation Offset

EL.3 100% Saturation mV value
EL.4 0% Saturation mV value
In mg/L (or ppm) mode:
EL.1 Electrode

Slope

Factor

EL.2 100% Saturation mV value
EL.3 0% Saturation mV value

SETUP

P 4.0

A 0

P4.0: Select Auto Power On/Off

P4.0

Select auto power on or off

SETUP

P 5.0

P5.0: Reset to Factory Default

P5.1

Reset calibrated data

P5.2

User reset back to factory default values

MEM

SETUP

P 6.0

P6.0: Memory Clear

P6.0 Clear

all

stored

readings