8.6.2

User Reset

This program clears all data including calibration,
memory and other customised setup functions and reset
it to factory default.

You can skip program P5.1 (Calibration Reset) and
proceed straight to Program P5.2:

1. From Program P5.0 - rSt, press HOLD/ENTER

key three times to P5.2. Press HOLD/ENTER key
again to enter the User Reset menu.

2. Press the MI/▲ or MR/▼ keys to select or

deselect the user reset function.

• ‘NO’ retains current settings

• ‘YES’ resets to factory default settings.

3. Press the ENTER key to confirm selection.

4. If ‘NO’ is selected, it will retains current settings

and goes back to program 5.0.

5. If ‘YES’ is selected, all LCD segments will light up

for one second, reset to factory default settings,
and the meter will be in the measurement mode.

6. Otherwise press CAL/MEAS key to return to

measurement mode without resetting to factory
default.

yes

SETUP

SETUP

SETUP

P 5.2

ENTER

HOLD

MI

MR

Select either "YES" or "nO",

press ENTER key to confirm.

Figure 23: P5.2 - User reset