6 MEMORY

FUNCTION

The DO 110 meter can store up to 100 sets of data. You can store data in any
combination of values such as “% Saturation and temperature” or “mg/L (or ppm) and
temperature”.

6.1

Memory Input

1. During any measurement function (MEAS), press MI/▲

key to input any data into the memory.

2. MEM, “Sto” and memory number will flash. The meter

then returns to measurement mode.

3. If the memory is full, the display will flash “MEM – FUL”

momentarily (Figure 7). The first value stored will be
erased to create space for the new value.

6.2

Memory Recall

This function recalls the previous readings stored in the memory. Memory recall is in
“Last In First Out” order. To recall readings:

1. Press the MR/▼ key once to retrieve the last reading stored. The memory

location screen – MEM, “Loc” and the memory number – will flash on the
display.

2. Press the HOLD/ENTER key to recall the reading stored

under that memory number.

3. To display next stored reading, press HOLD/ENTER key. The

display automatically moves to the next lower memory
location screen.

4. To view a stored reading in a specific memory location, press

MI/▲ or MR/▼ keys to locate it, and press
HOLD/ENTER key to display the reading.

NOTE: Readings stored in memory are retained even if the
unit is turned off. To erase all readings stored in memory, use the SETUP
in section 8.7 – P 6.0: Memory Clear (CLr) on page 31.