Page 32

background image

Instruction Manual

CyberScan DO 110

27

8.4.2

EL.2 Viewing the % Saturation Offset (Only available in % Saturation
Set Up menu):

This parameter shows you the amount of the % Saturation Offset entered in the other
parameter “OFS” P7.0.

From Step 3 above,

4. Press the HOLD/ENTER key to enter the %

Saturation Offset viewing menu. The meter will
display the last offset adjustment made.

8.4.3

Viewing the 100% Saturation mV value

This parameter shows the sensor’s mV output
corresponding to 100% Saturation.

From Step 4 above,

5. Press the HOLD/ENTER key. The display will show

the sensor’s mV output value with respect to 100%
Saturation.

8.4.4

Viewing the 0% Saturation mV value

This parameter lets you view the sensor’s mV output
corresponding to 0% Saturation.

From Step 5 above,

6. Press the HOLD/ENTER key. The display will show

the sensor’s mV output value with respect to 0%
Saturation.

7. Press the HOLD/ENTER key again to exit to the

subgroup menu. Press CAL to return to the
measurement mode.

Note: To view the electrode properties in the mg/L
(ppm), repeat the whole steps above using the mg/L (ppm) Set Up menu.
Viewing of the % Saturation Offset will be skipped in this menu and the
viewing sequence will as follows:

EL.1

Electrode Slope Factor

EL.2 100% Saturation mV value

EL.3 0% Saturation mV value

%

mV

mV

SETUP

SETUP

SETUP

EL.2

EL.3

EL.4

10.0

30.0

0.3

ENTER

HOLD

ENTER

HOLD

ENTER

HOLD

Figure 20: P3.0 - View % Saturation
offset, electrode mV value in 100%
and 0% saturation