El.2 el.3 el.4 – Thermo Fisher Scientific CyberScan DO 110 User Manual
Page 32
Instruction Manual
CyberScan DO 110
27
8.4.2
EL.2 Viewing the % Saturation Offset (Only available in % Saturation
Set Up menu):
This parameter shows you the amount of the % Saturation Offset entered in the other
parameter “OFS” P7.0.
From Step 3 above,
4. Press the HOLD/ENTER key to enter the %
Saturation Offset viewing menu. The meter will
display the last offset adjustment made.
8.4.3
Viewing the 100% Saturation mV value
This parameter shows the sensor’s mV output
corresponding to 100% Saturation.
From Step 4 above,
5. Press the HOLD/ENTER key. The display will show
the sensor’s mV output value with respect to 100%
Saturation.
8.4.4
Viewing the 0% Saturation mV value
This parameter lets you view the sensor’s mV output
corresponding to 0% Saturation.
From Step 5 above,
6. Press the HOLD/ENTER key. The display will show
the sensor’s mV output value with respect to 0%
Saturation.
7. Press the HOLD/ENTER key again to exit to the
subgroup menu. Press CAL to return to the
measurement mode.
Note: To view the electrode properties in the mg/L
(ppm), repeat the whole steps above using the mg/L (ppm) Set Up menu.
Viewing of the % Saturation Offset will be skipped in this menu and the
viewing sequence will as follows:
EL.1
Electrode Slope Factor
EL.2 100% Saturation mV value
EL.3 0% Saturation mV value
%
mV
mV
SETUP
SETUP
SETUP
EL.2
EL.3
EL.4
10.0
30.0
0.3
ENTER
HOLD
ENTER
HOLD
ENTER
HOLD
Figure 20: P3.0 - View % Saturation
offset, electrode mV value in 100%
and 0% saturation