8.2.2

P1.2: Select °C or °F temperature units

The DO 110 meter lets you select between °C and °F
units for temperature readings.

From measurement mode

1. Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P1.0.

3. Press the HOLD/ENTER key two times to select

parameter 1.2.

4. Press the MI/▲ or MR/▼ keys to toggle

between °C and °F.

5. Press the HOLD/ENTER key to confirm

selection and to return to the subgroup
menu. Press the CAL/MEAS key to return to
measurement mode.

8.2.3

P1.3: Selection of Automatic or Manual Temperature Compensation

This feature lets you select between Automatic
Temperature Compensation (ATC) and Manual
Temperature Compensation (MTC). Meter default is ATC.

From measurement mode

1

Press SETUP key to enter Set Up menu.

2 Press the HOLD/ENTER keys until you view

parameter P1.3.

3 Press the HOLD/ENTER key again. The upper

display shows “ATC” and the lower display shows
“YES” or “NO”.

4 Press

MI/▲ or MR/▼ keys to select between

Automatic and Manual Temperature
Compensation.

• “YES” activates the Automatic Temperature

Compensation function.

• “NO” activates the Manual Temperature

Compensation function.

5 Press the HOLD/ENTER key to confirm

selection and to return to the subgroup menu.

6 Press the CAL/MEAS key to return to

Figure 15: P1.2 - Select

temperature units, °C or °F

°F

°C

SETUP

f

p 1.2

SETUP

C

p 1.2

MR

MI

SETUP

YES

°C

ATC

SETUP

P 1.3

°C

ATC

SETUP

°C

HOLD

ENTER

MI

MR

Figure 16: P1.3-Select Automatic or
Manual Temperature Compensation