INSERT/DELETE A STEP


Insert a new step
• Touch Insert a new step to insert the next step in

the current stage.

• Edit the new step temperature, hold time and ramp

rate as required.

Delete a step

• To delete a step, touch the step area to highlight it

and then touch Delete.





INSERT/DELETE A PAUSE

Insert a pause
• To insert a pause between stages, touch Insert a

pause.


The hold temperature will be that of the previous step.
The unit will not continue with the remainder of the
program until it is manually un-paused by the user.

Delete a pause

• To delete a pause, touch the pause button to

highlight it and then touch Delete.

Note: To insert a pause between two previously
programmed stages, it is necessary to first delete the
later stages to where the pause is required.