6. Release the heated lid by lifting the latch forward

and use it to lift the lid. To close the lid, lower the lid

onto the block and press the latch to secure in

place.

7. Place the samples in the block. If individual tubes or

strip tubes are being used, space these out evenly

across the block to equalise the pressure from the

heated lid.


8. The heated lid has a rotating knob to adjust the lid

pressure on the samples, allowing for a variety of

consumables to be used. To adjust the pressure:

a. Place the samples in the block and place the lid

in the down position.

b. Adjust the orange rotating knob anticlockwise to

give the least pressure then close the lid fully.

c. Gently rotate the knob clockwise until you can

feel pressure beginning to be applied.

d. Rotate the knob a further quarter of a turn and

this will give the correct pressure. Do not over-

tighten.

9. Once the pressure has been set it does not need to be adjusted unless a different consumable is

used.