Techne PrimeG User Manual
Page 29

28
To increase the step hold time per cycle:
• For an increase in hold time per cycle, touch the
button below Increase Per Cycle to turn this
function on. Next touch the time entry button below
this.
A new screen will open which is similar to the simple
time entry screen. The maximum increase in hold
time per cycle (based on the number of cycles and
the time limits of the unit) will be shown.
• Enter the required value using the + and - buttons.
• If an entered incremented time is invalid, a prompt
will indicate the acceptable range.
• Touch OK to return to the previous screen.
The increase in hold time per cycle will now be
displayed.
• Touch OK again to return to the programming
screen.
• Touch Cancel to return to the programming screen
if no changes are required.
The time value will now appear orange in the
programming screen indicating that it contains a
modified function.