beautypg.com

Techne PS139 User Manual

Ps139

background image

Page 1 of 3

PS139 Issue 4 07/2014

PS139

Fitting stacking plates and wall plates on a Hybrigene


A trained technician should carry out this work only after reading the full instructions.

There are two kits available for stacking and wall fixing either two or three Hybrigene units:


FHB4DSTK: Stacking and wall mounting kit for two units

Description

Part code

Quantity

Wall Bracket

6105214

1

M4 x 8 Pozi Pan Screw

6008652

9

M4 Crinkle Washers

6008639

9

Thumbscrew

6009552

1

Wall Plate

6105213

1

Stacking Plate

6105212

1



FHB4DKIT
: Stacking and wall mounting kit for three units

Description

Part code

Quantity

Wall Bracket

6105214

1

M4 x 8 Pozi Pan Screw

6008652

15

M4 Crinkle Washers

6008639

15

Thumbscrew

6009552

1

Wall Plate

6105213

1

Stacking Plate

6105212

2


Fixing Procedure for two Hybrigenes


1. Ensure the units are turned off and remove the power to them.

2. Place one unit on the top of the other.

3. The stacking plate has three holes and three slots. It has three slots to allow for the

adjustment of the feet for levelling the units.


4. Use three of the screws to attach the plate to the bottom of the top unit.

5. Use another three screws to attach the plate to the bottom hybridiser.

6. Ensure that the units are level and the screws are tight.