Fitting stacking plates and wall plates on a Hybrigene
A trained technician should carry out this work only after reading the full instructions.
There are two kits available for stacking and wall fixing either two or three Hybrigene units:
FHB4DSTK: Stacking and wall mounting kit for two units
Description
Part code
Quantity
Wall Bracket
6105214
1
M4 x 8 Pozi Pan Screw
6008652
9
M4 Crinkle Washers
6008639
9
Thumbscrew
6009552
1
Wall Plate
6105213
1
Stacking Plate
6105212
1
FHB4DKIT: Stacking and wall mounting kit for three units
Description
Part code
Quantity
Wall Bracket
6105214
1
M4 x 8 Pozi Pan Screw
6008652
15
M4 Crinkle Washers
6008639
15
Thumbscrew
6009552
1
Wall Plate
6105213
1
Stacking Plate
6105212
2
Fixing Procedure for two Hybrigenes
1. Ensure the units are turned off and remove the power to them.
2. Place one unit on the top of the other.
3. The stacking plate has three holes and three slots. It has three slots to allow for the
adjustment of the feet for levelling the units.
4. Use three of the screws to attach the plate to the bottom of the top unit.
5. Use another three screws to attach the plate to the bottom hybridiser.
6. Ensure that the units are level and the screws are tight.