Initial Denaturation
Use this to denature the sample and/or activate the

enzyme before the thermal cycling program begins.

• Touch On/Off to toggle between settings.


• Touch the time button to set the required hold time.

• Touch the temperature button to set the required

temperature.


Hot Start
This will cause the unit to pause at the set

temperature. The program will not continue until the
unit is manually un-paused.

• Touch On/Off to toggle between settings.


• Touch the temperature button to set the required

temperature.

• Touch More Settings for further program defaults

or Back to return to the Settings menu.

Final Extension
Use this to add a final extension step to the program to
ensure complete amplification of the products.

• Touch On/Off to toggle between settings.

• Touch the time button to set the required hold time.

• Touch the temperature button to set the required

temperature.

Final Hold

This will hold the block at the set temperature at the
end of a program until the user manually stops the
unit.

• Touch On/Off to toggle between settings.

• Touch the temperature button to set the required

temperature.

• Touch More Settings to return to the top of the

Program Defaults menu or Back to return to the
Settings menu.

SERVICE SETTINGS


The service settings are for the use of authorised service engineers only. A service engineer’s
password is required to access these settings.