beautypg.com

Techne PrimeG User Manual

Page 22

background image

21

The Ice Bucket and Ligation programs are simple temperature holds, with the former having an infinite

hold at 10ºC and the latter at 15ºC. In both of these programs the heated lid is switched off and the

initial denaturation, hot start and final extension functions disabled.

INCREMENTED/DECREMENTED TIME AND TEMPERATURE

The software allows the user to either increment or decrement the time and/or temperature of a step

within a cycling stage. These features are used with applications such as touchdown PCR where the

annealing temperature is gradually decreased during the cycling process and long PCR, where due to

the size of the product, long annealing/extension times are required and the extension time is

increased by 15 to 20s per cycle during the final stages of the reaction.

GRADIENT (PRIMEG ONLY)

The gradient feature of the PrimeG can be useful in optimising the annealing conditions for primers. A

gradient can be set around a temperature in any step of a program. The set temperature is the

temperature in the middle columns of the block and the gradient is the variation at the two extremes of

the block; the left hand columns being the coolest and the right hand columns the hottest.

The maximum temperature gradient range which can be set is 29ºC

(depending on block type)

and the

minimum is 1ºC; the lowest temperature is 30ºC and the hottest temperature is 80ºC. Examples are

given in the table below:

Lowest temperature

(ºC)

Set temperature

(ºC)

Highest temperature

(ºC)

Gradient

(ºC)

50

55

60

10

40.5

55

69.5

29

50.5

65

79.5

29

A gradient calculator is included in the Settings module so the user can obtain an accurate estimation
of the temperature of each column during the gradient stage. The temperatures for any particular

gradient are calculated from an algorithm stored in the block firmware.

A
B
C
D
E
F
G
H

1 2 3 4 5 6 7 8 9 10 11 12

Set

temperature

Lowest

temperature

Highest

temperature

Gradient

A
B
C
D
E
F
G
H

1 2 3 4 5 6 7 8 9 10 11 12

A
B
C
D
E
F
G
H

1 2 3 4 5 6 7 8 9 10 11 12

Set

temperature

Lowest

temperature

Highest

temperature

Gradient

This manual is related to the following products: