Pressure Module Specifications
Range
Resolution
Accuracy
0-.3 psig
.0001
.03%FS± .07%RDG ± 1 digit
0-1 psig
.0001
.03%FS± .07%RDG ± 1 digit
0-5 psig
.001
.02%FS± .05%RDG ± 1 digit
0-10 psig
.001
.02%FS± .05%RDG ± 1 digit
0-15 psig
.001
.02%FS± .05%RDG ± 1 digit
0-30 psig
.001
.02%FS± .05%RDG ± 1 digit
0-50 psig
.001
.02%FS± .05%RDG ± 1 digit
0-100 psig
.01
.02%FS± .05%RDG ± 1 digit
0-300 psig
.01
.03%FS± .06%RDG ± 1 digit
0-1000 psig
.1
.03%FS± .06%RDG ± 1 digit
0-1500 psig
.1
.03%FS± .07%RDG ± 1 digit
0-3000 psig
.1
.1%FS
0-5000 psig
.1
.1%FS
0-1 psig vacuum
.0001
.03%FS± .07%RDG ± 1 digit
0-5 psig vacuum
.001
.02%FS± .05%RDG ± 1 digit
0-15 psig vacuum .001
.02%FS± .05%RDG ± 1 digit
0-15 psia
.001
.02%FS± .05%RDG ± 1 digit
0-30 psia
.001
.02%FS± .05%RDG ± 1 digit
0-50 psia
.001
.02%FS± .05%RDG ± 1 digit
Special ranges can be provided upon request.
Over Pressure
Operating, without damage or calibration shift:
0-100 psi, vac & abs
300% of sensor range
300-2000 psi
200% of sensor range
3000-5000 psi
150% of sensor range
A
±FS Plus ± %RDG ± 1 digit
Specifications
130/135