beautypg.com

Calibration of mv to current converters – Martel Electronics 135 User Manual

Page 58

background image

Calibration of mV to Current Converters


This example assumes a 10 to 50 mV input and a 4 to 20 mA output.

1.

Connect as shown in Figure 3.


2.

Select mA input.


3.

Select mV output.


4.

Enter 10 mV and adjust converter zero for 4 mA output.


5.

Enter 50 mV and adjust converter span for 20 mA output.


6.

Repeat steps 4 and 5 until both zero and span are within desired
tolerance.


















Figure 3

Application Notes

130/135

54








mV/Current

Converter

OUTPUT


- +





INPUT


- +

Input 10-50 mV

Output 4-20 mA

This manual is related to the following products: