beautypg.com

Martel Electronics 135 User Manual

Page 62

background image

No external module present

You pressed the OPT key and there is no external module attached to the
calibrator. If there is a module attached, turn the calibrator off and back on and
attempt the operation again. If the error persists the module is probably
defective.

No mode change in MENU/SETUP

You have pressed the INPUT or OUTPUT key while a menu is active or in Setup
mode. You must complete the menu or setup sequence before changing modes.
You may cancel a menu by pressing CE or exit from Setup by pressing SETUP a
second time.

Output section out of tolerance

The output section of the calibrator is out of calibration tolerance to the extent
that the calibrator is not usable. Power the calibrator off and back on to reset the
error. If the error persists the calibrator must be serviced.

Output section fatal error det

The output section of the calibrator has malfunctioned to the extent that
operation is not possible. Power the calibrator off and back on to reset the error.
If the error persists the calibrator must be serviced.

Ramp not available for this function

The selected output function does not support Ramp mode.

Unidentified err CalledBy ??????

An internal error has been detected by the calibrator firmware. Turn the
calibrator off and back on to reset the error. If the problem persists, note the
data in the lower right of the display and contact your service representative.

Wrong access code entered

You entered the wrong access code when attempting to change the owner’s
name or accessing the manufacturing test menu. The access code for changing
the owner name is found on the factory calibration sheet. The access code for
accessing the test menu is found in the BetaCal documentation for users who are
performing their own calibrations.




Error Messages

130/135

58

This manual is related to the following products: