Service Information

130/135

61

Fuse Type/

Location

Fuse

No.

Function

Test

Input/
Output

Model 130

Model 135

Top PCB
130/135



Model 130
Model 135



F2

F3

mA Output Short

the

milliamps

output

terminals and attempt to output
10mA. If the output reads Open,
the fuse may be blown. Measure
from the mA positive (red) terminal
to the volts out negative (black)
terminal. If you measure 13-15 V
the fuse is OK.


F3
F4

V/mV

Output

If mA and frequency output
function normally, but volts or
millivolts do not function, the fuse
may be blown.

Main Power


Bottom
PCB
130/135

F1

Primary

Power

This fuse should not blow unless an
internal malfunction occurs or
unless the user inadvertently shorts
the main DC power line while the
case is open. The symptom would
be that the unit is completely dead.

F2

Charger

This fuse protects the batteries in
case

the

wrong

charger

is

connected to the unit. The
symptom would be that the unit
would not recharge from the
charger. If this fuse is found blown,
ensure that you are using the
charger supplied with the calibrator.