Error Messages

Entered value is out of range

The value entered is outside of the allowable range for the current output
function. Check the specifications for the function being used to determine the
allowable range of values.

Expecting ENTER CE or INC/DEC

The only keystrokes that are acceptable at this time are the ENTER, CE, INC or
DEC.

Expecting ENTER or CE/CANCEL

The only keystrokes that are acceptable at this time are the ENTER or CE
(interpreted as Cancel).

Illegal key for current function

The active function does not recognize the last key pressed. Review the
instructions for the function you are using.

INPUT function not expecting data

You are attempting to enter numeric data and the keyboard is in input mode. The
keyboard should be in output mode (press the OUTPUT key) before entering
numeric data.

Input section out of tolerance

The input section of the calibrator is out of calibration tolerance to the extent that
the calibrator is not usable. Power the calibrator off and back on to reset the
error. If the error persists the calibrator must be serviced.

Input section fatal error det

The input section of the calibrator has malfunctioned to the extent that the
operation is not possible. Power the calibrator off and back on to reset the error.
If the error persists the calibrator must be serviced.

KEY not implemented

You have pressed a key that has no meaning.