OPERATION

4

25

OPTIWAVE 5200 C

www.krohne.com

09/2013 - 4002500101 - AD FF OPTIWAVE 5200 R01 en

FD_FAIL_MASK

FD_FAIL_MASK

FD_FAIL_MASK

FD_FAIL_MASK

Fail Mask

Fail Mask

Fail Mask

Fail Mask

RW

This parameter allows the user to suppress any single or

multiple conditions that are active, in this category, from

being broadcast to the host through the alarm parameter.

(0)-(31) masked

Setting:

(00) Check Function

(01) Cond01

(02) Cond02

(03) Cond03

(04) Cond04

(05) Cond05

(06) Cond06 disabled

(07) Bad Measurement quality disabled

(08) Signal Strong disabled

(09) Signal Weak disabled

(10) Empty Spectrum invalid disabled

(11) Sensor No Signal

(12) Sensor Microwave Error disabled

(13) Temperature Above +75°C disabled

(14) Temperature Below -35°C disabled

(15) Overfill Warning disabled

(16) Level Lost warning disabled

(17) Temperature out of range (warning) disabled

(18) Local operation disabled

(19) Overfill Error disabled

(20) Level lost error disabled

(21) Temperature out of range Error disabled

(22) Converter Voltage Error disabled

(23) Converter ROM Error disabled

(24) Converter RAM Error disabled

(25) Converter EEPROM Error disabled

(26) Sensor not compatible disabled

(27) Internal Communication Error disabled

(28) Sensor Voltage Error disabled

(29) Sensor ROM Error disabled

(30) Sensor RAM Error disabled

(31) Sensor EEPROM Error disabled

FD_FAIL_PRI

FD_FAIL_PRI

FD_FAIL_PRI

FD_FAIL_PRI

Fail Priority

Fail Priority

Fail Priority

Fail Priority

RW

This parameter allows the user to specify the priority of this

alarm category.

0

Setting:

0...15

FD_MAINT_ACTIVE

FD_MAINT_ACTIVE

FD_MAINT_ACTIVE

FD_MAINT_ACTIVE

Maintenance Active

Maintenance Active

Maintenance Active

Maintenance Active

R

This parameter reflects the error conditions that are being

detected as active as selected for the "Maintenance" category.

It is a bit string, so that multiple conditions may be shown.

FD_MAINT_ALM

FD_MAINT_ALM

FD_MAINT_ALM

FD_MAINT_ALM

Fail Maintenance

Fail Maintenance

Fail Maintenance

Fail Maintenance

Alarm

Alarm

Alarm

Alarm

This parameter is used primarily to broadcast a change in the

associated active conditions, which are not masked, for this

alarm category to a Host System.

UNACKNOWLEDGED

Unacknowledged

RW

A discrete enumeration which is set to "Unacknowledged"

when an alarm occurs, and set to "Acknowledged" by a write

from a human interface device or other entity which can

acknowledge that the alarm/event has been noticed.

Uninitialized

Setting:

Uninitialized / Acknowledged / Unacknowledged

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value