beautypg.com

Operation – KROHNE OPTIWAVE 5200 FOUNDATION FIELDBUS User Manual

Page 23

background image

OPERATION

4

23

OPTIWAVE 5200 C

www.krohne.com

09/2013 - 4002500101 - AD FF OPTIWAVE 5200 R01 en

FD_FAIL_ACTIVE

FD_FAIL_ACTIVE

FD_FAIL_ACTIVE

FD_FAIL_ACTIVE

Fail Active

Fail Active

Fail Active

Fail Active

R

This parameter reflects the error conditions that are being

detected as active as selected for this category. It is a bit

string, so that multiple conditions may be shown.

Uninitialized

Setting:

(00) Check Function

(01) Cond01

(02) Cond02

(03) Cond03

(04) Cond04

(05) Cond05

(06) Cond06

(07) Bad Measurement quality

(08) Signal Strong

(09) Signal Weak

(10) Empty Spectrum invalid

(11) Sensor No Signal

(12) Sensor Microwave Error

(13) Temperatur Above +75°C

(14) Temperature Below -35°C

(15) Overfill Warning

(16) Level Lost warning

(17) Temperature out of range (warning)

(18) Local operation

(19) Overfill Error

(20) Level lost error

(21) Temperature out of range Error

(22) Converter Voltage Error

(23) Converter ROM Error

(24) Converter RAM Error

(25) Converter EEPROM Error

(26) Sensor not compatible

(27) Internal Communication Error

(28) Sensor Voltage Error

(29) Sensor ROM Error

(30) Sensor RAM Error

(31) Sensor EEPROM Error

FD_FAIL_ALM

FD_FAIL_ALM

FD_FAIL_ALM

FD_FAIL_ALM

Fail Diagnostic Alarm

Fail Diagnostic Alarm

Fail Diagnostic Alarm

Fail Diagnostic Alarm

This parameter is used primarily to broadcast a change in the

associated active conditions, which are not masked, for this

alarm category to a Host System.

UNACKNOWLEDGED

Unacknowledged

RW

A discrete enumeration which is set to "Unacknowledged"

when an alarm occurs, and set to "Acknowledged" by a write

from a human interface device or other entity which can

acknowledge that the alarm/event has been noticed.

Uninitialized

Setting:

Uninitialized / Acknowledged / Unacknowledged

ALARM_STATE

Alarm State

R

A discrete enumeration which gives an indication of whether

the alert is active and whether it has been reported.

Uninitialized

Setting:

Uninitialized / Clear - reported / Clear - not reported / Active -

reported / Active - not reported

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value