beautypg.com

KROHNE IFC 010 Converter EN User Manual

Page 61

background image

10/4

Power supply

1. Rated voltage

tolerance band

2. Rated voltage

tolerance band

Frequency

Power consumption
(incl. primary head)

Housing
Material

polycarbonate (PC)

Protection category (IEC 529/EN 60 529)
IFC 010 K (compact)

IP 67, equivalent to NEMA 6, same as primary head

IFC 010 F (separate)

IP 65, equivalent to NEMA 4 / 4X

1. AC-Version

2. AC-Version

3. AC-Version

DC-Version

Standard

Option

Option

Option

230 / 240 V

200 V

48 V

24 V

200 – 260 V

170 – 220 V

41 – 53 V

11 – 32 V

115 / 120 V

100 V

24 V

100 – 130 V

85 – 110 V

20 – 26 V

48 – 63 Hz

approx. 5 VA

approx. 4.5 W

When connected to a functional extra-low voltage, 11-32 V DC, protective
separation (PELV) must be ensured (VDE 0100/VDE 0106 and IEC 364/IEC 536)

dia. 9

(0.35”)

27.5

(1.08”)

1 output cable (see Sect. 2.3)
2 signal cable of primary head

(see Sect. 1.3)

3 field power cable of primary head

(see Sect. 1.3)

4 power supply cable (see Sect. 1.2)
5 internal connection

(ssee Fig. in Sect. 8.9, plug connectors
X3 and X5)

1

16 (4.57”)

160 (6.30”)

144 (5.67”)

9 (0.35”)

dia. 9

(0.35”)

4

5

258 (10.16”)

220 (8.66”)

1

2

3

4

34 (1.34”)

120 (4.72”)

1

1

IFC 010 F and ZD dimensions and weights 10.4

ZD Intermediate
connection box
Weight approx.
0.5 kg/1.1 lbs

IFC 010 F
weight approx. 3.8 kg / 8.4 lb

dimensions in mm

PG 16

1

/

2

” NPT

57 (4.02”)

22 (4.02”)

80

125 (4.92”)

68 (2.68”)

(3.15”)