7/5

Part 2

Signal converter IFC 010 D (D = display version) and
Signal converter IFC 010 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.1)

Group D

Display shows . . .

Cause

Remedial action

D 1

LINE INT.

Power failure.

Delete error message in

Note: no counting during

RESET/QUIT. menu. Reset

power failure.

totalizer(s) if need be.

D 2

CUR.OUTP. I

Current output overranged.

Check instrument parameters
and correct if necessary.
Error message deleted
automatically after cause
has been eliminated.

D 3

PULS.OUTP. P

Pulse output overranged.

Check instrument parameters,

Note: totalizer deviation

correct if necessary, and reset

possible

totalizer(s). Error message
deleted automatically
after cause has been
eliminated.

D 4

ADC

Analog/digital converter

Error message deleted

overranged.

automatically after cause
has been eliminated.

D 5

FATAL. ERROR

Fatal Error,

Replace signal converter

all outputs set to “min.”

(see Sect. 8.4) or contact

values.

Krohne Service, having
first noted down hardware
information and error status,
see Fct. 2.02.

D 6

TOTALIZER

Counts lost

Delete error message in

(overflow, data error)

RESET/QUIT. menu.

D 7

„STARTUP“

Hardware fault,

Replace signal converter

cyclic flashing

Watchdog activated.

(see Sect. 8.4) or contact
Krohne Service.

D 8

BUSY

Displays for flow,

Change setting in

totalizers and errors disabled.

Fct. 1.04.

D 9

Unsteady display

– Process product

Increase time constant,

conductivity too low,

see Sect. 6.1, or contact

particles/air inclusions

Krohne Service.

too large or inhomogeneous

– Pulsating flow
– Time constant too low

D 10

No display

Power OFF.

Switch on power

Check power fuse(s)

Replace if defective

F1 (F1 and F2 with DC).

(see Sect. 8.2).

Group I

Fault / Symptom

Cause

Remedial action

I 1

Receiver instrument

Incorrect connection/polarity

Connect properly,

indicates “0”.

see Sect. 2.3.

Receiver instrument or current

Check output (see Sect. 7.2)

output defective.

with new milliammeter:
Test ok, check connection
cables and receiver instrument,
replace if necessary.
Test faulty, current output
defective. Replace signal
converter (see Sect. 8.4)
or contact Krohne Service

Current output disabled,

Activate under Fct. 1.05.

see Fct. 1.05

Short-circuit between current

Check connection and cables,

output and pulse output.

see Sect. 2.3. Voltage between
I+ and I

approx. 15 V.

Switch off device, eliminate
short-circuit, and switch
device on again.

I 2

Unsteady display

– Process product

Increase time constant,

conductivity too low,

see Sect. 6.1, or contact

particles/air inclusions

Krohne Service.

too large or inhomogeneous

– Pulsating flow
– Time constant too low