KROHNE IFC 010 Converter EN User Manual
Page 46
Always switch off power source before opening the housing !
Required measuring instruments and tools
-
A crosstip screwdriver
-
Ohmmeter with at least 6 V range
or AC voltage/resistance measuring bridge.
Note: Accurate measurements in the area of the electrodes can only be obtained with an
AC voltage/resistance bridge. Also, the measured resistance is very heavily dependent on
the electrical conductivity of the liquid product.
Preparatory work
•
Switch off power source.
•
Remove transparent cover
(unscrew 4 recessed head screws)
and black plastic cover
(unscrew 1 recessed head screw),
see Fig. A and B in Sect. 8.5.
•
Detach blue 9-pin plug,
see Fig. D in Sect. 8.5,
field power supply (pins 7+ 8)
and signal cables (pins 1, 2, 3, 4 + 5).
•
Completely fill the measuring tube of the
flowmeter with the process liquid.
7.5 Test of primary head
Jacks X1 and X4 on amplifier PCB,
see Sect. 8.9
Blue 9-pin plug
(connection to primary head)
unassigned
X1
X4
7 8 1 2 3 4 5
7/7
Action
Typical result
Incorrect result =
Resistance measurement
flowmeter defective,
at the blue 9-pin plug
return to factory for repair,
refer to last but one page !
1
Measure resistance
30 - 150
Ω
if lower:
between wires
interwinding fault
7 and 8.
if higher:
wire break
2
Measure
> 10 M
Ω
if lower:
resistance between
interwinding fault to PE
U-clamp terminal
or FE.
in terminal compartment
(PE protective conductor
or FE functional ground)
and wires 7 and 8
3
Measure resistance
1 k
Ω
- 1 M
Ω
if lower: drain measuring
between wires 1 and 3
(see “Note” above).
tube and repeat
and between 1 and 4
Both values should be
measurement; if still
(same measuring lead
approx. equal.
too low, short-circuit in
always on wire 1 ! )
electrode wires.
if higher: break in
electrode wires or
electrodes
contaminated.
Values not equal:
break in electrode
wires or electrodes
contaminated.