Always switch off power source before opening the housing !

Required measuring instruments and tools
-

A crosstip screwdriver

-

Ohmmeter with at least 6 V range

or AC voltage/resistance measuring bridge.
Note: Accurate measurements in the area of the electrodes can only be obtained with an
AC voltage/resistance bridge. Also, the measured resistance is very heavily dependent on
the electrical conductivity of the liquid product.

Preparatory work

Switch off power source.

Remove transparent cover
(unscrew 4 recessed head screws)
and black plastic cover
(unscrew 1 recessed head screw),
see Fig. A and B in Sect. 8.5.

Detach blue 9-pin plug,
see Fig. D in Sect. 8.5,
field power supply (pins 7+ 8)
and signal cables (pins 1, 2, 3, 4 + 5).

Completely fill the measuring tube of the
flowmeter with the process liquid.

7.5 Test of primary head

Jacks X1 and X4 on amplifier PCB,

see Sect. 8.9

Blue 9-pin plug

(connection to primary head)

unassigned

X1

X4

7 8 1 2 3 4 5

7/7

Action

Typical result

Incorrect result =

Resistance measurement

flowmeter defective,

at the blue 9-pin plug

return to factory for repair,
refer to last but one page !

1

Measure resistance

30 - 150

if lower:

between wires

interwinding fault

7 and 8.

if higher:
wire break

2

Measure

> 10 M

if lower:

resistance between

interwinding fault to PE

U-clamp terminal

or FE.

in terminal compartment
(PE protective conductor
or FE functional ground)
and wires 7 and 8

3

Measure resistance

1 k

- 1 M

if lower: drain measuring

between wires 1 and 3

(see “Note” above).

tube and repeat

and between 1 and 4

Both values should be

measurement; if still

(same measuring lead

approx. equal.

too low, short-circuit in

always on wire 1 ! )

electrode wires.

if higher: break in
electrode wires or
electrodes
contaminated.

Values not equal:
break in electrode
wires or electrodes
contaminated.