4.5 Error messages in measuring mode
The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES“ set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.“.
Fct.
Text
Description and settings
3.05
USER UNIT
Set any required unit for flowrate and counting
→
TEXT VOL.
Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press
↵
key to transfer to subfunction “FACT. VOL.”
→
FACT. VOL.
Set conversion factor (F
M
) for volume
Factory set “1.00000” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1
x
10
3
or 2.64172
x
10
-4
).
Factor F
M
= volume per 1m
3
.
Setting range
• 1.00000 E-9 to 9.99999 E+9 (= 10
-9
to 10
+9
)
Press
↵
key to transfer to subfunction “TEXT TIME”.
→
TEXT TIME
Set text for required time unit (max. 3 characters)
Factory-set: “hr” or “day”:
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press
↵
key to transfer to subfunction “FACT. TIME”
→
FACT. TIME
Set conversion factor (F
T
) for time
Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6
x
10
3
or 8.64
x
10
-4
).
Set factor F
T
in seconds.
Setting range
• 1.00000 E-9 to 9.99999 E+9 (= 10
-9
to 10
+9
)
Press
↵
key to return to Fct. 3.05 “USER UNIT”.
3.06
APPLICAT.
Set overload point for A/D converter
→
EMPTY PIPE
Switch on “empty tube” identifier option?
(appears only when this option is installed)
• YES • NO
Select with key
↑
or
↓
.
Press
↵
key to return to
Fct. 3.06 “APPLICAT.“.
Error messages
Description of error
Error clearance
LINE INT.
Power failure Note:
Cancel error in RESET/QUIT. menu
no counting during power failure
Reset totalizer if necessary.
CUR. OUTP. I
Current output overranged.
Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.
PULSOUTP. P
Pulse output overranged.
Check and if necessary correct
Note:
instrument parameters. After
totalizer deviation possible.
elimination of cause, error message
deleted automatically.
ADC
Analog / digital
Error message deleted automatically
converter overranged
after elimination of cause.
FATAL. ERROR
Fatal error, all outputs set
Please consult factory.
to “min. values“
TOTALIZER
Totalizer has been reset
Cancel error message in
RESET/QUIT. menu.
EMPTY PIPE
Pipe has run dry.
Fill pipe.
This message appears only when
the “empty pipe identifier”
option is installed and the function
is switched on under Fct. 3.06
APPLICAT., submenu
“EMPTY PIPE”.