KROHNE IFC 010 Converter EN User Manual

Page 24

background image

4/9

4.5 Error messages in measuring mode

The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES“ set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.“.

Fct.

Text

Description and settings

3.05

USER UNIT

Set any required unit for flowrate and counting

TEXT VOL.

Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).

Press

key to transfer to subfunction “FACT. VOL.”

FACT. VOL.

Set conversion factor (F

M

) for volume

Factory set “1.00000” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1

x

10

3

or 2.64172

x

10

-4

).

Factor F

M

= volume per 1m

3

.

Setting range

1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to transfer to subfunction “TEXT TIME”.

TEXT TIME

Set text for required time unit (max. 3 characters)
Factory-set: “hr” or “day”:
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).

Press

key to transfer to subfunction “FACT. TIME”

FACT. TIME

Set conversion factor (F

T

) for time

Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6

x

10

3

or 8.64

x

10

-4

).

Set factor F

T

in seconds.

Setting range
1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to return to Fct. 3.05 “USER UNIT”.

3.06

APPLICAT.

Set overload point for A/D converter

EMPTY PIPE

Switch on “empty tube” identifier option?
(appears only when this option is installed)
YES NO

Select with key

or

.

Press

key to return to

Fct. 3.06 “APPLICAT.“.

Error messages

Description of error

Error clearance

LINE INT.

Power failure Note:

Cancel error in RESET/QUIT. menu

no counting during power failure

Reset totalizer if necessary.

CUR. OUTP. I

Current output overranged.

Check and if necessary correct
instrument parameters. After
elimination of cause, error message
deleted automatically.

PULSOUTP. P

Pulse output overranged.

Check and if necessary correct

Note:

instrument parameters. After

totalizer deviation possible.

elimination of cause, error message
deleted automatically.

ADC

Analog / digital

Error message deleted automatically

converter overranged

after elimination of cause.

FATAL. ERROR

Fatal error, all outputs set

Please consult factory.

to “min. values“

TOTALIZER

Totalizer has been reset

Cancel error message in
RESET/QUIT. menu.

EMPTY PIPE

Pipe has run dry.

Fill pipe.

This message appears only when
the “empty pipe identifier”
option is installed
and the function
is switched on under Fct. 3.06
APPLICAT., submenu
“EMPTY PIPE”.