7.7 Test of signal converter using GS 8 A simulator (option)

GS 8 A operating elements and accessories

+
0

A

5-to-3 pin adapter
for cable C5

B

Plug for field power
supply, 2-pin

C3 / C5

Plug for signal cable,
3-pin / 5-pin

D

Switch,
flow direction

H

Socket for H2 plug
on cable Z

H2

Plug of cable Z

L

Power supply ON

P

Potentiometer “zero”

X3

Socket for plug B
on amplifier PCB

X5

Socket for plug C3
on amplifier PCB

Y

Switch,
measuring ranges

Z

Cable between GS 8 A
and signal converter

Electrical connection

H2

Z

C 5

B

X 5

X 3

amplifier PCB, see
Sect. 8.9

Connection of milliammeter and
electronic frequency counter,
see Sect. 2.3 “Connection of outputs”.

10

Power

Outputs

U-clamp terminal

I

I I+ P P

S

PE
FE
FE

Milliammeter, accuracy
class 0.1, R

i

< 500 ohms,

range 4-20 mA
Electronic frequency
counter, input resistance
approx. 1 kohm, range
0-1 kHz, time base min. 1
second, see connection
diagrams in Sect. 2.3.2.

A

a) Switch off power source before

opening the housing !

b) Unscrew the 4 recessed head screws, see

Fig. A in Sect. 8.5, and remove transparent
cover from the signal converter housing.

c) Unscrew recessed head screw, see Fig. B in

Sect. 8.5, and remove black plastic cover.

d) Pull off blue 7-pin plug from the amplifier PCB,

see Sect. 8.9:
socket X3 field power supply and socket X5
signal cables.

e) Connect plug B to socket X3 (2-pin) and

plug C (5-pin) to socket X5 (3-pin) by way
of adapter A (5-to-3pin).

1,25

2,5

5

10

20

320

160

80

40

20

Y

P

D

L

H

L

N

1L~ 0L~
L+ L-

Using GS 8 simulator

Additional adapter is required

between GS 8 and IFC 010 signal converter.

Order No. 2.10764.00.

Warning: Instrument must be properly grounded to avoid personnel shock hazard.