KROHNE IFC 010 Converter EN User Manual
Page 25
4/10
Reset totalizer and cancel error messages, RESET / QUIT menu 4.6
Cancel error messages in RESET / QUIT menu
Key
Display
Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
CodE 2
- -
Key in entry code 2 for RESET / QUIT
menu:
↑ →
↑ →
ERROR QUIT.
Menu for error acknowledgement
→
QUIT. NO
Do not delete error messages,
press
↵
twice = return to measuring mode
↑
QUIT. YES
Delete error messages
↵
ERROR QUIT.
Error messages deleted
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
Reset totalizer(s) in RESET / QUIT menu
Key
Display
Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
CodE 2
- -
Key in entry code 2 for RESET / QUIT
menu:
↑ →
↑ →
ERROR QUIT.
Menu for error acknowledgement
↑
TOTAL. RESET
Menu for resetting totalizer
→
RESET NO
Do not reset totalizer,
press
↵
twice = return to measuring mode
↑
RESET. YES
Reset totalizer
↵
RESET QUIT.
Totalizer reset
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
Examples of setting the signal converter 4.7
The cursor, flashing part of display, is shown below in bold type.
•
Change measuring range of current output and value for error messages (Fct. 1.05):
•
Change measuring range from 04-20 mA to 00 -20 mA
•
Change value for error messages from 0 mA to 22 mA
Key
Display
Description
→
If “YES” set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now:
→ → → ↵ ↵ ↵ ↑ ↑ ↑
Fct. 1.00
OPERATION
→
Fct. 1.01
FULL SCALE
4x
↑
Fct. 1.05
CURRENT I
→
FUNCT. I
→ ↵
RANGE I
If “REV. RANGE” appears here,
press keys
→
and
↵
again.
→
04-20
mA
Old current range
↑
00-20
mA
New current range
↵
I ERROR
→
0
mA
Old value for error messages
2x
↑
22
mA
New value for error messages
↵
Fct. 1.05
CURRENT I
↵
Fct. 1.00
OPERATION
↵
STORE YES
↵
- - - - - - -
- - - - - / - - -
Measuring range with new data for the current output