Reset totalizer and cancel error messages, RESET / QUIT menu 4.6

Cancel error messages in RESET / QUIT menu

Key

Display

Description

- - - - - - -

- - - - - / - - -

Measuring mode

CodE 2

- -

Key in entry code 2 for RESET / QUIT
menu:

↑ →

↑ →

ERROR QUIT.

Menu for error acknowledgement

QUIT. NO

Do not delete error messages,
press

twice = return to measuring mode

QUIT. YES

Delete error messages

ERROR QUIT.

Error messages deleted

- - - - - - -

- - - - - / - - -

Return to measuring mode

Reset totalizer(s) in RESET / QUIT menu

Key

Display

Description

- - - - - - -

- - - - - / - - -

Measuring mode

CodE 2

- -

Key in entry code 2 for RESET / QUIT
menu:

↑ →

↑ →

ERROR QUIT.

Menu for error acknowledgement

TOTAL. RESET

Menu for resetting totalizer

RESET NO

Do not reset totalizer,
press

twice = return to measuring mode

RESET. YES

Reset totalizer

RESET QUIT.

Totalizer reset

- - - - - - -

- - - - - / - - -

Return to measuring mode

Examples of setting the signal converter 4.7

The cursor, flashing part of display, is shown below in bold type.

Change measuring range of current output and value for error messages (Fct. 1.05):

Change measuring range from 04-20 mA to 00 -20 mA

Change value for error messages from 0 mA to 22 mA

Key

Display

Description

If “YES” set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

Fct. 1.00

OPERATION

Fct. 1.01

FULL SCALE

4x

Fct. 1.05

CURRENT I

FUNCT. I

→ ↵

RANGE I

If “REV. RANGE” appears here,
press keys

and

again.

04-20

mA

Old current range

00-20

mA

New current range

I ERROR

0

mA

Old value for error messages

2x

22

mA

New value for error messages

Fct. 1.05

CURRENT I

Fct. 1.00

OPERATION

STORE YES

- - - - - - -

- - - - - / - - -

Measuring range with new data for the current output