KROHNE IFC 010 Converter EN User Manual
Page 60
10/3
10.3 IFC 010 Signal converter
Versions
B - version
without display / control unit (basic version)
D - version
with display / control unit
Add-on equipment (option)
- CONFIG software and RS 232 adapter for control via
MS-DOS-PC, connection to IMoCom interface
- Hand-Held-Terminal for control of basic versions
- Other bus and computer interfaces on request
Current output
Function
- all operating data settable, galvanically isolated
Current
0 - 20 mA and 4 - 20 mA
Active output
load max. 500 ohms
Passive output
external voltage:
15 ... 20 V DC
20 ... 32 V DC
load: min. ... max.
0 ... 500
Ω
250 ... 750
Ω
Error identification
0 / 3.6 / 22 mA
Forward/reverse measurement
direction identified via status output
Pulse output
Function
- all operating data settable
- galvanically isolated
- digital pulse division, interpulse period non-uniform,
therefore if frequency meters connected allow for minimum
counting interval:
gate time, totalizer
≥
Pulse rate for Q = 100 %
10, 100 or 1000 pulses per second (= Hz), fixed or optionally
adjustable in pulses per liter, m
3
or US gallons
(special version: up to 10 kHz scaling)
Active output:
connection: electronic totalizer (EC)
internal voltage: approx. 15 V DC, from current output
load rating: I
max
< 23 mA when operated without current output
load rating:
I
max
<
2
3 mA when operated with current output
Passive output:
connection electromechanical (EMC) or electronic (EC)
totalizers
external voltage: U
ext
≤
30 V DC /
≤
24 V AC
load current: I
max
≤
150 mA
Pulse width
50, 100, 200, 500 ms or 1 s,
selectable with frequencies below 10 Hz
Forward/reverse measurement
direction identified via status output
Status output (passive)
Function
settable as indicator for flow direction, errors or trip point
Connection
external voltage: U
ext.
≤
30 V DC /
≤
24 V AC
load current: I
max
≤
150 mA
Time constant
0.2 to 99.9 seconds, settable in increments of 0.1 second
Low-flow cutoff
cutoff ”on” value: 1 to 19 % of Q
100%
, adjustable in
cutoff ”off” value: 2 to 20 % 1 % increments
Local display (D versions only)
3-line LCD
Display functions
actual flowrate, forward, reverse and sum totalizers (7-digit)
or 25-character bar graph with percent display and
status messages
Display units:
actual flowrate
settable in liter/s, m
3
/h, US gallons/min or user-defined unit,
e.g. hectoliter/day or US million gallons/day
totalizers
liter, m
3
or US gallons and 1 user-defined unit
(e.g. hectoliter), selectable overflow time
Language of plain texts
English, German, French, others on request
Display:
1st (top) line
8-character, 7-segment numeral and sign display,
symbols for key acknowledgement
2nd (middle) line
10-character, 14-segment text display
3rd (bottom) line
6 markers
t
to identify display
}
1000
P
100%
[Hz]