Error limits at reference conditions 10.2

Pulse output
± F
Error in % of flowrate (actual value):

Curve A: DN 10 – 600 /

3

/

8

”- 24”

v

0.4 m/s or v

1.3 ft/s : ± 0.5 % of measured value

v

0.4 m/s or v

1.3 ft/s : ± 0.002 m/s or 0.0066 ft/s

Curve B: DN 700 – 1000 / 28” – 40”

v

0.25 m/s or v

0.8 ft/s : ± 0.8 % of measured value

v

0.25 m/s or v

0.8 ft/s : ± 0.002 m/s or 0.0066 ft/s

Q Actual flowrate
Q

F

Flow for error limit v

F

= 0.25 or 0.8 ft/s (see Flow tables)

v

Flow velocity in m/s and ft/s

v

F

Flow velocity in m/s and ft/s at Q

F

(see Flow tables)

Reference conditions
Product

Water, 10 to 30°C / 50 to 86°F

Electrical conductivity

> 300

µ

S/cm (

µ

mho/cm)

Power supply (line voltage)

U

N

(± 2%)

Ambient temperature

20 to 22°C / 68 to 71.6°F

Warm-up time

30 minutes

Straight inlet run

> 10

x

DN (DN = meter size)

Straight outlet run

> 3

x

DN

Primary heads

properly grounded and centered

Current output

same as above error limit for pulse output plus ...
0 to 20 mA:

± 0.05 % of full-scale range in each case

4 to 20 mA:

± 0.062 %

}

}

± F 3

%

2

1

0.8
0.5

0.25 0.4 0.5 0.75 1 10 11 12
(0.8) (1.3) (1.6) (2.4) (3.2) (33) (36) (40)

Q

F

Q

F

B
A

v

ms

(ft / sec)

}