KROHNE IFC 010 Converter EN User Manual
Page 42
7/3
7.4 Faults and symptoms during start-up and process flow measurement
Before contacting Krohne Service, please read through the
following tables. THANK YOU !
•
Most faults and symptoms occurring with the compact flowmeters can be eliminated
with the aid of the following tables.
•
For greater clarity, faults and symptoms in the tables are divided into 2 parts and
various groups.
•
Part 1
Signal converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG user program (see Sect. 6.1)
Groups:
LED
LED display (status messages)
I
Current output
P
Pulse output
LED / I / P LED display, current output and pulse output
•
Part 2
Signal converter IFC 010 D (D = display version) and
Signal converter IFC 010 B (B = basic version), without display
but with CONFIG user program (see Sect. 6.1)
Groups:
D
Display
I
Current output
P
Pulse output
S
Status indication output
D / I / P / S LED display, current output,
status output
Part 1
Converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG operator program
Group LED
Fault / Symptom
Cause
Remedial action
LED 1
LED flashes red/green
Overranging of A/D converter,
Reduce flowrate; if unsuccessful,
current or pulse output
test as described in Sect. 7.5.
Measuring tube drained,
Fill measuring tube.
A/D converter overranged.
LED 2
LED flashes red
Fatal Error, hardware and/or
Replace signal converter
software error
(see Sect. 8.4) or contact
Krohne Service.
LED 3
cyclic flashing of red LED,
Hardware fault, Watchdog
Replace signal converter
approx 1 second
activated.
(see Sect. 8.4) or contact
Krohne Service.
LED 4
LED shows red continuously
Hardware fault
Replace signal converter
(see Sect. 8.4) or contact
Krohne Service.