KROHNE IFC 010 Converter EN User Manual

Page 14

background image

3/1

3 Start-up

Before powering the system, please check that it has been correctly installed according
to Sect. 1 and 2.

The flowmeter is delivered ready for operational use. All operating data have
been factory set in accordance with your specifications.
Please refer to Sect. 3.2 “factory settings”.

Power the unit, and the flowmeter will immediately start process flow
measurement.

Basic version, signal converter IFC 010 / B

A light emitting diode (LED) under the cover of the electronic section shows the
measurement status.
LED flashing . . .

green:

measurement correct, everything all right.

green / red: momentary overdriving of outputs

and/or A/D converter.

red:

fatal error, parameter error or hardware fault,
please consult factory.

Refer to Sect. 6.1 for operator control of the “basic version”.

Display version, signal converter IFC 010 / D

When powered, the display shows in succession: START UP and READY.
This is followed by display of the current flow rate and/or the current totalizer count
on either a continuous or alternating basis, depending on the setting under Fct. 1.04.

Refer to Sect. 4 and 5 for operator control of the “display version”.

3.1 Power- on and measurement