KROHNE IFC 010 Converter EN User Manual

Page 23

background image

4/8

Fct.

Text

Description and settings

3.00

INSTALL.

Installation menu

3.01

LANGUAGE

Select language for display texts
GB / USA (English)

F (French)

D (German)

others on request

Press

key to return to Fct. 3.01

“LANGUAGE”.

3.02

FLOWMETER

Set data for primary head

DIAMETER

Select size from meter size table
DN 10 - 1000 mm equivalent to 3/8 - 40 inch

Select with

or

key.

Press

key to transfer to subfunction “FULL SCALE ”.

FULL SCALE

Full-scale range for flow Q

100%

To set, refer to Fct. 1.01 “FULL SCALE” above.

Press

key to transfer to subfunction “GKL VALUE”.

VALUE P

Pulse value (Fct. 1.06

“VALUE P”

) has been changed.

With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

GKL VALUE

Set primary constant GKL
see primary head nameplate.
Range:

1.0000 - 9.9999

Press

key to transfer to subfunction

“FIELD. FREQ.”.

FIELD FREQ.

Magnetic field frequency
Values: 1/6 or 1/18 of power frequency, see nameplate.

Press

key to transfer to subfunction “FLOW DIR.”;

DC units only: to transfer to subfunction

“LINE FREQ.”.

LINE FREQ.

Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:

50 Hz

and

60 Hz

Press

key to transfer to subfunction “FLOW DIR.”.

FLOW DIR.

Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
+ DIR.

– DIR.

Select using

or

key.

Press

key to return to Fct. 3.02 “FLOWMETER ”.

3.03

ZERO SET

Zero calibration
Note:

Carry out only at “0” flow and with completely filled
measuring tube!

Precautionary query
CALIB. NO

Press

key to return to Fct. 3.03 “ZERO SET”.

CALIB. YES

Press

key to start calibration.

Duration approx. 15 –90 seconds, current flowrate
displayed in the selected unit (see Fct. 1.04 “DISP. FLOW”).

A “WARNING” sign appears when flowrate “>

0

“;

acknowledge by pressing

key.

STORE NO (do

not store new zero value)

STORE YES (store new zero value)

Press

key to return to Fct. 3.03 “ZERO SET”.

3.04

ENTRY CODE

Entry code required to enter setting mode?
NO

(= entry with

only)

YES (= entry with

and Code 1:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

)

Press

to return to Fct. 3.04 “ENTRY CODE”.