H250 M40

www.krohne.com

11/2012 - 4000640702 MA H250 M40 R02 en

H250U - Vertical installation position

Reference condition for gas measurements:

Reference condition for gas measurements:

Reference condition for gas measurements:

Reference condition for gas measurements:
Flow measurements for gases are attributed to
Nl/h or Nm

3

/h: Volume current in standard state 0°C - 1.013 bar abs. (DIN 1343)

Measuring span:

10 : 1

Flow values:

Values = 100%

Water 20°C / 68°F

Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]

Flow direction

vertical downwards

EN

ASME

Cone

Water [l/h]

Air [Nm

3

/h]

Pressure loss

[mbar]

DN15

½

"

K 15.1

65

1.6

175

K 15.2

110

2.5

178

K 15.3

170

4

180

K 15.4

260

6

200

K 15.5

420

10

220

K 15.6

650

16

290

K 15.7

1100

28

520

K 15.8

1500

40

840

DN25

1"

K 25.1

1150

30

97

K 25.2

1800

45

85

K 25.3

2700

70

92

K 25.4

4500

120

115

K 25.5

7600

200

172

DN50

2"

K 55.1

9000

240

220

K 55.2

15000

400

230

K 55.3

21000

550

240

INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.