OPERATION

52

H250 M40

www.krohne.com

11/2012 - 4000640702 MA H250 M40 R02 en

FUNCTION B3

3.6

INACTIVE

The binary switching input B3 has no function.

ACTIV H

The internal flow counter is reset to 0.0 if there is an H-level at input B3

for at least 100 ms.

ACTIV L

The internal flow counter is reset to 0.0 if there is an L-level at input B3

for at least 100 ms.

STARTH STOPL

The counter is started by creating an H level at the B3 input and stopped

by creating an L level at the B3 input.

STARTL STOPH

The counter is started by creating an L level at the B3 input and stopped

by creating an H level at the B3 input.

MULTIDROP

3.7

0…15

Polling adress for HART

®

Multi-Drop Mode

When the address is 0, the HART

®

multi-drop mode is disabled.

Attention:

Attention:

Attention:

Attention:
When the HART

®

multi-drop mode (Address 1…15) is activated, the

current loop is inactive (fixed current value of 4.5 mA) and no longer

follows the current process.

4mA CALIBR.

3.8

D/A calibration for beginning of measuring range (4mA)

Note:

Note:

Note:

Note:
This function is not available in HART

®

multi-drop mode.

Attention:

Attention:

Attention:

Attention:

During calibration, the value of the current loop does not follow the

current process.

If a current measuring device in the 4...20mA loop detects a deviation

from the desired value 4,000mA, the measured value must be entered.

The correction value is accepted following the query Save and Confirm

with YES.

20mA CALIBR.

3.9

D/A calibration for the final value of the measuring range (20mA)

Note:

Note:

Note:

Note:
This function is not available in HART

®

multi-drop mode.

Attention:

Attention:

Attention:

Attention:

During calibration, the value of the current loop does not follow the

current process.

If a current measuring device in the 4...20mA loop detects a deviation

from the desired value 20,000mA, the measured value must be entered.

The correction value is accepted following the query Save and Confirm

with YES.

ALARM

CURRENT

3.10

OFF

Error indication via the current loop is deactivated. The current loop

follows the current process.

Note:

Note:

Note:

Note:
This function is not available in HART

®

multi-drop mode.

ALARM HIGH

Error indication via the current loop is activated (failure signal "high" as

per NE43).

Note:

Note:

Note:

Note:
This function is not available in HART

®

multi-drop mode.

ALARM LOW

Error indication via the current loop is activated (failure signal "low" as

per NE43).

Note:

Note:

Note:

Note:
1) This function is not available in HART

®

multi-drop mode.

2) This function is supported as of Electronic Revision 2.2.x.