6

OPERATION

42

H250 M40

www.krohne.com

11/2012 - 4000640702 MA H250 M40 R02 en

FRAM WRITE

FAULT

Internal

communication error

Error

Check whether the display is plugged in

correctly and restart the device. If the error

occurs again: send the device back to the

manufacturer with an indication of the error.

ROM/FLASH

ERROR

Memory error

detected during self-

test.

Error

Restart device. If the error occurs again:

send the device back to the manufacturer

with an indication of the error.

RESTART OF

DEVICE

A device restart has

taken place

information

The device has been restarted using menu

item 1.6.2 since the last time the error

messages were reset.

MULTIDROP

MODE

The HART® multidrop

mode is activated.

The current output is

set to a fixed value of

4.5 mA.

information

The HART

®

multidrop mode is activated with

selection of a polling address not equal to 0

using menu item 3.7. . The polling address 0

reactivates the analog output.

CRYSTAL OSC

FAULT

Internal error in

device

Error

The device must be sent back to the

manufacturer with an indication of the error.

REF VOLTAGE

FAULT

Internal error in

device

SENSOR A FAULT Internal error in

device

SENSOR B

FAULT

Internal error in

device

MEMORY

CORRUPTION

Internal memory

error, caused by a

hardware or software

problem

Error

Restart the device; if the error occurs again

the device must be sent back to the

manufacturer with an indication of the error.

AO FIXED

The current output is

set to a fixed value.

information

The current output is fixed and does not

reflect the measured value. This is the case

in multidrop mode, with current output
test/calibration using the menu or HART

®

.

AO SATURATED

Current output

saturated

information

The current output is saturated at 20.4 or

22.0 mA (depending on whether the alarm

current is activated or deactivated in menu

item 3.10), and is no longer coupled with the

measured value.

ERROR TIMEOUT

Data not transferred,

or transferred

incorrectly from the

ESK to the counter

module

Error

Confirm menu point 1.6.3 WRITE INFO I/O.

WARNING

TIMEOUT

Warning

Error Message

Description

Category

Remedy