Operation – KROHNE H250 M40 EN User Manual

Page 47

background image

OPERATION

6

47

H250 M40

www.krohne.com

11/2012 - 4000640702 MA H250 M40 R02 en

OUTPUT B2

1.2

Output B2 is a binary switching output. Under Fct. 3.4, one of the

following functions can be attributed to this output

INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.

One of the following functions can be selected as contact type under Fct. 3.5:

NO contact 1 / NC contact 2

1.2.1

INACTIVE

FLOW.VAL B2

see FL.R. VALUE B1

Under Fct. 1.2.2 a hysteresis can also be specified.

COUNTER.VAL B2

See COUNTER.VAL B1

PULSE WDTH B2

see PULSE WDTH Fct. 1.1.1

Configuration takes place under Fct. 1.3.1 (pulse width) and Fct. 1.3.2

(pulse/unit).

1.2.2

HYST.B2

See HYST. B1

PULSEOUTP.

1.3

1.3.1

Pulse width

50ms

T

i

= 50 ms; f

max

= 10 Hz

max. pulse / h = 36000

4ms

T

i

= 100 ms; f

max

= 5 Hz

max. pulse / h = 18000

200ms

T

i

= 200 ms; f

max

= 2,5 Hz

max. pulse / h = 9000

500ms

T

i

= 500 ms; f

max

= 1 Hz

max. pulse / h = 3600

1.3.2

PULSE/UNIT

0.001 ...

1000

Pulses per volume or mass unit, which can be output via one of the

binary outputs.

The maximum frequency for the pulse output (see Fct 1.3.1) cannot be

exceeded, even at maximum flow rate (final value).

Example:

Example:

Example:

Example:

Final value Qmax = 1200 l/h; volume unit counter = Litre; pulse

duration = 100 ms;

If the factor 1 is entered, 1 pulse/litre = 1200 pulses are generated in

one hour at maximum flow rate.

Number of max. permissible pulses:

Label

Level Select /

Input

Explanation

MA_H250_M40_R02_en_640702_PRT.book Page 47 Wednesday, November 28, 2012 12:23 PM