Technical data – KROHNE H250 M40 EN User Manual
Page 73
TECHNICAL DATA
8
73
H250 M40
www.krohne.com
11/2012 - 4000640702 MA H250 M40 R02 en
H250/C - Ceramic/PTFE
Reference condition during gas measurements:
Reference condition during gas measurements:
Reference condition during gas measurements:
Reference condition during gas measurements:
Flow measurements for gases are attributed to
Nl/h or Nm
3
/h: Volume current in standard state 0°C - 1.013 bar abs. (DIN 1343)
Measuring span:
10 : 1
Flow values:
Values = 100%
Water 20°C / 68°F
Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
Flow
Max. pressure loss
Water
Air
Water
Air
Liner /
Float
PTFE
Ceramic
PTFE
Ceramic
PTFE
Ceramic
PTFE
Ceramic
Nominal
meter size
Cone
[l/h]
[Nm
3
/h]
[mbar]
DN15, ½"
E 17.2
25
30
0,7
-
65
62
65
62
E 17.3
40
50
1,1
1.8
66
64
66
64
E 17.4
63
70
1,8
2.4
66
66
66
66
E 17.5
100
130
2,8
4
68
68
68
68
E 17.6
160
200
4,8
6.5
72
70
72
70
E 17.7
250
250
7
9
86
72
86
72
E 17.8
400
-
10
-
111
-
111
-
DN25, 1"
E 27.1
630
500
16
18
70
55
70
55
E 27.2
1000
700
30
22
80
60
80
60
E 27.3
1600
1100
45
30
108
70
108
70
E 27.4
2500
1600
70
50
158
82
158
82
E 27.5
4000
1
2500
120
75
290
100
194
100
DN50, 2"
E 57.1
4000
4500
110
140
81
70
81
70
E 57.2
6300
6300
180
200
110
80
110
80
E 57.3
10000
11000
250
350
170
110
170
110
E 57.4 16000
1
-
-
-
284
-
-
-
DN80, 3"
E 87.1
16000
16000
-
-
81
70
-
-
E 87.2
25000
25000
-
-
95
85
-
-
E 87.3 40000
1
-
-
-
243
-
-
-
DN100, 4"
E 107.1
40000
-
-
-
100
-
-
-
E 107.2 60000
1
-
-
-
225
-
-
-
1 special float
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.