Technical data – KROHNE OPTISENS MAC 080 EN User Manual
Page 43
TECHNICAL DATA
8
43
OPTISENS MAC 080
www.krohne.com
01/2010 - MA MAC 080 R02 en
Module for 2 additional
outputs (option)
Only one optional module can be installed. Either the 2 additional outputs module or
the Profibus DP module.
Functions: to extend the converter analog outputs to a total of 4.
Design: PCB only, will be installed inside of the converter box.
Ambient temperature: -20...+50°C / -4...+122°F, if installed in a operational converter
Storage temperature: 0…+60°C / +32…+140°F
Dimensions: 86 x 54 mm / 3.39" x 2.13"
Weight: 35 g / 1.23 oz
Output signals
Output signals
Output signals
Output signals
Two 4…20 mA outputs
Galvanic isolated
Load: 450 Ω
Profibus DP module (option)
Only one optional module can be installed. Either the 2 additional outputs module or
the Profibus DP module.
Functions: for connection of converter to a Profibus network as a slave and to enable
connection of four sensors.
Interface type: DP-V0 slave communication according to EN 50170 (DIN 19245,
IEC 61158), protocol version 1.10
Design: PCB only, will be installed inside of the converter box.
Ambient temperature: -20...+50°C / -4...+122°F, if installed in a operational converter
Storage temperature: 0…+70°C / +32…+158°F
Dimensions: 86 x 54 mm / 3.39" x 2.13"
Power: +5 V, max. 350 mA
Output: RS485 optically isolated Profibus interface with DC/DC converter on board;
Automatic baudrate detection (96000 bit/s...12 Mbit/s)
Physical interface
Physical interface
Physical interface
Physical interface
Transmission media: Profibus bus line, type A or B specified in EN 50170
Topology: Master-Slave communication
Profibus connectors: screw terminal
Cable: shielded copper cable, twisted pair
Isolation: the bus is galvanically separated from the other electronics with an on board
DC/DC converter. Bus signals (A-line and B-line) are isolated via optocouplers.
Configuration and indications
Configuration and indications
Configuration and indications
Configuration and indications
Address range: 1...99 using switches on the module or 1...126 via converter menus
Bus termination: switch inboard
LED indications: ON-line, OFF-line, Profibus related diagnostic
Approvals
Approvals
CE
This device fulfils the statutory requirements of the EC directives.
The manufacturer certifies successful testing of the product by applying the CE
marking.
EMC
Electromagnetic compatibility (EMC) in accordance with:
EN 61000-6-4:2001 Emission standard for industrial environments
EN 61000-6-2:2001 Immunity for industrial environments
Low Voltage Directive
Safety requirements for electrical equipment for measurement, control, and laboratory
use in accordance with EN 61010-1:2001
.book Page 43 Tuesday, January 19, 2010 1:07 PM