8

TECHNICAL DATA

42

OPTISENS MAC 080

www.krohne.com

01/2010 - MA MAC 080 R02 en

Installation conditions

Installation

Wall or handrail mountable (using optional available mounting plate)

Dimensions and weights

For detailed information refer to "Dimensions and weights".

Materials

Enclosure

Polycarbonate

Cable feedthrough

Nickel plated brass; inserts: polyamid

Solenoid valve (wetted parts)

Enclosure: brass

Internal parts: Stainless Steel

Gaskets: EPDM

Electrical connections

Power supply

85…250 VAC; 50/60 Hz

Power consumption

20 W (0.18 A at 110 V)

Fuse

3.15 A slow 250 V, 4 x 20 mm

Connection cable

5-pin M12 contact; fixed on sensor side, shielded; 10 m / 33 ft long

Cable feedthrough

Standard: M20 x 1.5 (IP68 rated)

Option: ½" NPT adapter

Inputs and outputs

Sensor inputs (up to four)

Digital communication RS485

One direct sensor input via M12 plug

2...4 via external junction box (option)

Cable length: 10 m / 33 ft (max. 100 m / 328 ft)

Digital inputs

3 digital inputs (for selection of different calibration curves)

Analog outputs

Two 4…20 mA outputs, (extendable to four via expansion module)

Galvanic isolated

Load: 450 Ω

Relay outputs

2 relays

Dry contacts, normally open

Max. 250 VAC, max. 12 A

Serial interface

RS232 "log line" to PC (mainly used for service)

19200 Baud, 8 bits, 1 stop bit, no parity, no flow control

Using simple ASCII protocol described in the manual.

Profibus (option)

Profibus DP-V0 (slave) module