ELECTRICAL CONNECTIONS

4

21

OPTISENS MAC 080

www.krohne.com

01/2010 - MA MAC 080 R02 en

4.3.1 Analog outputs

The two analog outputs of the converter are default active, sourcing 4...20 mA into a load of
maximum 450 Ω. They are galvanic isolated from the rest of the system, but the two channels
use a common ground.

Channel 2 can be jumpered to be passive, and fully isolated sink 4...20 mA from an external
supply of max. 24 VDC, by changing JP1 and JP7.

Analog output channel 2

Position JP1

Position JP7

Active, sourcing (default)

1-3 and 2-4

1-2

Passive, sinking

1-2 and 3-4

3-4

Figure 4-3: Active analog output (channel 1 or channel 2)

1 External load passive
2 Converter terminal 6 or 8
3 Converter active sourcing 4...20 mA
4 Converter terminal 7 or 9

Figure 4-4: Passive analog output (channel 2 only)

1 External load
2 External 24 VDC
3 Converter terminal 8
4 Converter passive sinking 4...20 mA (channel 2 only)
5 Converter terminal 9

.book Page 21 Tuesday, January 19, 2010 1:07 PM