Operation – KROHNE OPTISENS MAC 080 EN User Manual
Page 35
OPERATION
6
35
OPTISENS MAC 080
www.krohne.com
01/2010 - MA MAC 080 R02 en
6.2.2 Menu "System"
Menu
Description
Version
Version
Version
Version
Program version, read only.
Serial
Serial
Serial
Serial
Circuit board S/N, read only.
Box temp
Box temp
Box temp
Box temp
Internal box temperature, read only.
Exp.Module
Exp.Module
Exp.Module
Exp.Module
Press ^ to see the Fieldbus status sub menu.
Fieldbus
Yes or no, read only.
Installed
Yes or no, read only.
Hardwarefail
Yes or no, read only.
Initialised
Yes or no, read only.
Online
Yes or no, read only.
Software v.
Software version of the Profibus module, read only.
Out data
Last data sent to master, read only.
Err Cmd
Last error, read only
Err Cmd 1
Last error, read only
SW Reset
Restart the fieldbus module.
Test
Test
Test
Test
Press ^ to go to the system test sub menu.
Dig.in
Calibration curve selected by the digital inputs.
Relay1
"Off" / "On"; Be careful if something is connected to the relay. If the
relay is configured for a function, it cannot be changed.
Relay2
"Off" / "On"; Be careful if something is connected to the relay. If the
relay is configured for a function, it cannot be changed.
Analog 1 mA
Analog output 1, press ^ to get 4/20 mA output.
Analog 2 mA
Analog output 2, press ^ to get 4/20 mA output.
Analog 3 mA
Analog output 3 (only with extra 4...20 mA module).
Analog 4 mA
Analog output 4 (only with extra 4...20 mA module).
Panel LED
Green / Red.
Box heat
Off / On, read only.
Service
Service
Service
Service
Locked service menu for internal use of manufacturer.
.book Page 35 Tuesday, January 19, 2010 1:07 PM