PROFIBUS SETTINGS

5

35

IFC 300

www.krohne.com

02/2011 - 4001086601 - AD IFC 300 PROFIBUS R01 en

5.3 Menu C, setup

No.

Function

Settings / descriptions

C3 I/O totalizer

C3.1

FB2 totalizer 1

Set function of totalizer …

…

stands for 1, 2, 3 (= totalizer 1, 2, 3)

C3.2

FB3 totalizer 2

C3.3

FB4 totalizer 3

C3.….1

funct. of totalizer

Select:

sum totalizer (counts positive and negative values) /

+totalizer (counts only the positive values) /

-totalizer (counts only the negative values) /

hold totalizer (totalizer is stopped, no counting) /

all as positive (neg. input will be multiplied with -1.0) /

all as negative (pos. values will be multiplied with 1.0)

C3.….2

measurement

Selection of the measurement for totalizer …

Select: volume flow / mass flow (not

not

not

not valid for PF (partly filled))

C3.….3

preset value

Predefines a threshold (using high low limit value of the totalizer affected;

the THRESHOLD bit will be set in the long status information bytes of the

interface if the actual value of the totalizer is outside these limits. This can

be also used for a status output.

C3.….4

reset totalizer

The current value of the totalizer can be set to zero.

Select: no / yes (reset totalizer 1…3)

C3.….5

error behaviour

Defines the behaviour of this function block in case of errors.

Select: hold meas. value / ignore error / stop totalizer

hold meas. value:

hold meas. value:

hold meas. value:

hold meas. value: Totalization is continued based on the last incoming value

with good status before the first occurrence of bad status.

ignore error:

ignore error:

ignore error:

ignore error: Totalization is continued using the input values despite the bad

status. The status is ignored.

stop totalizer:

stop totalizer:

stop totalizer:

stop totalizer: Totalization is stopped during occurrence of bad status of

incoming values.

C3.….6

information

Serial no. of the I/O board, software version no. and production date of the

circuit board will be displayed

C4 I/O PROFIBUS

C4

I/O PROFIBUS

Using the menu functions mentioned below you will be able to control

basically the five analog input blocks of this PROFIBUS device. These five

menus are identical so they are grouped together and their functions are

described in one go.

C4.1

FB1 analog inp.

There are 5 analog input blocks.

…

stands for the 5 analog input blocks: FB1 (… = 1), FB5 (… = 2), FB6 (… = 3),

FB7 (… = 4) and FB8 (… = 5)

C4.2

FB5 analog inp.

C4.3

FB6 analog inp.

C4.4

FB7 analog inp.

C4.5

FB8 analog inp.

C4.….1

measurement

Select measurement for the analog input blocks:

flow speed / volume flow / mass flow (not

not

not

not valid for PF (partly filled)) / coil

temperature / conductivity (not

not

not

not valid for PF (partly filled) and CAP

(capacitive)) / temperature (electronic temperature) / supply (internal supply

voltage for PROFIBUS interface)

C4.….2

time constant

Set time constant for this function block (…).