Application profile – KROHNE FC 300 Converter PROFIBUS PA-DP User Manual

Page 30

background image

4

APPLICATION PROFILE

30

IFC 300

www.krohne.com

02/2011 - 4001086601 - AD IFC 300 PROFIBUS R01 en

4.3.3 DIAGNOSIS_EXTENSION (if "Classic Status" and "Diagnosis" selected)

Octet numb

er

Bit number

Subpar

ameter

Comment*

1

0

ADC_ELEC_TEMPERATURE

(S) electronic temperature

1

ADC_GAIN_ERROR

(S) gain error

2

ADC_F_FREQ_TOO_HIGH

(S) field frequency too high

3

Reserved

4

ADC_COIL_TEMPERATURE

(S) coil temperature

5

ADC_F_CURR_DEVIATION

(S) field current deviation

6

ADC_F_COIL_BRIDGED

(S) field coil bridged

7

ADC_F_COIL_BROKEN

(S) field coil broken

2

0

ADC_ELECTRODE_NOISE

(S) electrode noise

1

ADC_FLOW_PROFILE

(S) flow profile

2

ADC_LINEARITY

(S) linearity

3

ADC_ELECTRODE_SYM

(S) electrode symmetry

4

Reserved

5

Reserved

6

ADC_PIPE_NOT_FULL

(S) pipe not full

7

ADC_EMPTY_PIPE_0

(S) empty pipe (Uncertain)

3

0

Reserved

1

ADC_FIELD_FREQ_HIGH

(F) field frequency too high

2

ADC_EMPTY_PIPE_1

(F) empty pipe (Application Error)

3

ADC_DC_OFFSET

(F) DC Offset

4

Reserved

5

ADC_FIELD_CURR_LOCAL

(F) field current local

(F) field current local

(F) field current local

(F) field current local

6

ADC_SENSOR_LOCAL

(F) sensor local

(F) sensor local

(F) sensor local

(F) sensor local

7

ADC_SENSOR_GLOBAL

(F) sensor global

(F) sensor global

(F) sensor global

(F) sensor global

4

0

ADC_FLOW_EXCEED_LIM0

(F) flow exceeding limit

1

ADC_FLOW_EXCEED_LIM1

(F) flow exceeding limit

2

MEAS_AVAILABLE_NO

(I) measurement value not available

3

POWER ON ERROR

(I) power fail

4

ADC_TEST_SENSOR

(C) test sensor

5

ADC_WARNING_W1

(S) uncertain measurement

6

ADC_WARNING_W2

(F) application error

7

ADC_SENSOR_ELECTRONIC

(F) sensor electronic

(F) sensor electronic

(F) sensor electronic

(F) sensor electronic