Operation, 1 settable functions – KROHNE IFC 100 Converter FOUNDATION FIELDBUS User Manual
Page 9
OPERATION
4
9
IFC 100
www.krohne.com
02/2012 - 4001702901 - AD IFC 100 FF R01 en
4.1 Settable functions
• For Foundation Fieldbus the totalizers of the signal converter are not available!
• The following tables describe only the menus, functions and parameters that are different
between the standard signal converter and the Foundation Fieldbus signal converter.
• For the electrical connections of the outputs, inputs and all settings of functions that are not
listed in the following tables refer to the standard product documentation.
No.
Displayed text
Description and settings
A quick setup
A2
Tag
Identifier for the measurement in a plant, appears in the display header
(maximum 8 digits).
Note: Only read, not changeable!
B test
B3.5
Foundation fieldbus
Display information about Foundation fieldbus interface.
C setup
C1.2.3
time constant
For all flow measurements and outputs.
xxx.x s; range: 0.0…100 s
Note: Only read, not changeable!
C5.1.1
Tag
Identifier for the measurement in a plant, appears in the display header
(maximum 8 digits).
Note: Only read, not changeable!
C3
Totalizer
Not available for Foundation Fieldbus devices!
C4
I/O HART
Not available for Foundation Fieldbus devices!
C5.4
2. meas. page
In case of Foundation Fieldbus device, the second measurement page is
meant to check the output values of the different function blocks. Only
Foundation Fieldbus values can be selected here. The analogue inputs are
shown with exactly the value, seen on the bus system.
C5.4.1
measurement 1.line
Select: AI1 analog inp. / AI2 analog inp. / AI3 analog inp. / PID / INT1
integrator / INT2 integrator /
C5.4.2
format 1.line
Fixed number of digits after the decimal point or automatic, where the
number of digits is automatically adjusted to the available space.
C5.4.3
measurement 2.line
Select: AI1 analog inp. / AI2 analog inp. / AI3 analog inp. / PID / INT1
integrator / INT2 integrator /
C5.4.4
format 2.line
Fixed number of digits after the decimal point or automatic, where the
number of digits is automatically adjusted to the available space.
C5.4.5
measurement 3.line
Select: AI1 analog inp. / AI2 analog inp. / AI3 analog inp. / PID / INT1
integrator / INT2 integrator /
C5.4.6
format 3.line
Fixed number of digits after the decimal point or automatic, where the
number of digits is automatically adjusted to the available space.
C5.8
Foundation Field
-
C5.8.1
Simulate
Select: disabled / enabled
C5.8.2
Information
Display information about hardware and software versions, the calibration
and test date of this interface.