B.54 [source:]voltage:protection:over command, B.55 [source:]voltage:protection:over? query, B.58 [source:]voltage:protection:under command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.87 to 3.xx User Manual
Page 73: B.59 [source:]voltage:protection:under? query, B.53, Source, B.54, B.55, B.56, B.57

SERIES EL 022013
B-11
B.53 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? QUERY
VOLT?
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]?
Returns:
Description: Indicates voltage setpoint.
B.54 [SOURce:]VOLTage:PROTection:OVER COMMAND
VOLT:PROT:OVER
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVER
Long Form: [SOURce:]VOLTage:PROTection:OVER
where
Description: Sets overvoltage protection limit. Overvoltage protection setpoint is stored in non-volatile memory;
value is restored upon power-up.
B.55 [SOURce:]VOLTage:PROTection:OVER? QUERY
VOLT:PROT:OVER?
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVER?
Long Form: [SOURce:]VOLTage:PROTection:OVER?
Returns
Description: Indicates overvoltage protection limit.
B.56 [SOURce:]VOLTage:PROTection:OVER:STATe[:LEVel] COMMAND
VOLT:PROT:OVER:STAT 0
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVER:STAT[:LEV] 0
Long Form: [SOURce:]VOLTage:PROTection:OVER:STATe[:LEVel] 0
Description: Clears overvoltage protection error.
B.57 [SOURce:]VOLTage:PROTection:OVER:STATe[:LEVel]? COMMAND
VOLT:PROT:OVER:STAT?
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVER:STAT[:LEV]
Long Form: [SOURce:]VOLTage:PROTection:OVER:STATe[:LEVel]
where
0 (overvoltage protection error has not occurred)
Description: Indicates whether an overvoltage protection has occurred (1) or not (0).
B.58 [SOURce:]VOLTage:PROTection:UNDer COMMAND
VOLT:PROT:UND
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND
Long Form: [SOURce:]VOLTage:PROTection:UNDer
where
Description: Sets undervoltage protection limit. Undervoltage protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up.
B.59 [SOURce:]VOLTage:PROTection:UNDer? QUERY
VOLT:PROT:UND?
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND?
Long Form: [SOURce:]VOLTage:PROTection:UNDer?
Returns:
Description: Indicates undervoltage protection limit.