beautypg.com

B.5 diagnostic:tree:separator command, B.6 diagnostic:tree:separator? query, B.7 lock command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.87 to 3.xx User Manual

Page 65: B.8 lock? query, B.9 [source:]input[:state] command, B.10 [source:]input[:state]? query, B.11 [source:]input:cutoff:time command, Diagnostic:tree, Diagnostic:tree:sepa, Source

background image

SERIES EL 022013

B-3

B.5

DIAGnostic:TREE:SEParator COMMAND

DIAG:TREE:SEP

Syntax:

Short Form: DIAG:TREE:SEP

Long Form: DIAGnostic:TREE:SEParator

where defines the column separator

Description: Defines the column separator used when producing a text file of the command set using

DIAG:TREE:DISP command.

B.6

DIAGnostic:TREE:SEParator? QUERY

DIAG:TREE:SEP?

Syntax:

Short Form: DIAG:TREE:SEP?

Long Form: DIAGnostic:TREE:SEParator?

Returns defines the column separator

Description: Returns the number corresponding to the selected column separator when producing a text

file of the command set using DIAG:TREE:DISP command.

B.7

LOCK COMMAND

LOCK

Syntax:

Short Form: LOCK

Long Form: LOCK

where = OFF, EMERgency or LOCKed

Description: Controls functionality of front panel controls. When set to OFF, all front panel controls are fully

functional. When set to EMER, the front panel LOAD switch can only turn the load to the off state; all
other front panel controls except for the POWER switch are disabled. When set to LOCK, all front
panel controls except for the POWER switch are disabled.

B.8

LOCK? QUERY

LOCK?

Syntax:

Short Form: LOCK?

Long Form: LOCK?

Returns: = OFF, EMERgency or LOCKed

Description: Indicates state of front panel controls.

B.9

[SOURce:]INPut[:STATe] COMMAND

INP

Syntax:

Short Form: [SOUR:]INP[:STAT]

Long Form: [SOURce:]INPut[:STATe]

where = 0 or OFF (load disengaged) or 1 or ON (load engaged)

Description: Used to engage or disengage the load from the UUT. When set to 1 the load engages at the pro-

grammed mode and corresponding setpoint value; when set to 0 the load disengages.

B.10 [SOURce:]INPut[:STATe]? QUERY

INP?

Syntax:

Short Form: [SOUR:]INP[:STAT]?

Long Form: [SOURce:]INPut[:STATe]?

Returns = 0 or OFF (load disengaged), 1 or ON (load engaged), or 1,DIS (load disabled)

Description: Indicates whether load is engaged or disengaged from the UUT. When set to 1 the load is

engaged, when set to 0 the load is disengaged. When INP? returns 1,DIS it indicates that the output
has been disabled because input voltage was lost while the load was engaged and input voltage is
below the cutoff voltage established by INP:CUT:VOLT.

B.11 [SOURce:]INPut:CUToff:TIME COMMAND

INP:CUT:TIME

Syntax:

Short Form: [SOURce:]INP:CUT:TIME
Long Form: [SOURce:]INPut:CUToff:TIME
where = .0000 to 9999.000 (seconds)

Description: Establishes the amount of time to wait for input voltage to recover after input is in Disabled

state. Input disabled state occurs when input voltage falls below value established by INP:CUT:VOLT
while load was engaged (see PAR. B.13). While input is disabled, the LOAD indicator blinks green/
amber and INP? returns 1,DIS. If input is still disabled after time established by INP:CUT:TIM, the
input state is set to off: INP? returns 0 and the LOAD indicator shows steady green.This command is
only effective if INP:CUT:VOLT command is implemented.