B.25 [SOURce:]MEASure[:SCALar]:CURRent[:DC]? QUERY
MEAS:CURR?
Syntax:
Short Form: [SOUR:]MEAS[:SCAL]:CURR[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:CURRent[:DC]?
Returns: Measured values.
Description: Measures current. Measured value returned as XXX Amps, where XXX is decimal data representing
measured current.
B.26 [SOURce:]MEASure[:SCALar]:POWer[:DC]? QUERY
MEAS:POW?
Syntax:
Short Form: [SOUR:]MEAS[:SCAL]:POW[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:POWer[:DC]?
Returns: Measured values in the form 0.000 Kilowatts
Description: Measures power. Measured value returned as XXX Kilowatts, where XXX represents measured
power.
B.27 [SOURce:]MEASure[:SCALar]:VOLTage[:DC]? QUERY
MEAS:VOLT?
Syntax:
Short Form: [SOUR:]MEAS[:SCAL]:VOLT[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:VOLTage[:DC]?
Returns: Measured values in the form XXX Volts
Description: Measures voltage. Measured value returned as XXX Volts, where XXX is decimal data representing
measured voltage.
B.28 [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude] COMMAND
COND
Syntax:
Short Form:[SOUR:]COND[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]
Where
Description: Establishes conductance setpoint. Conductance setpoint affects the load only when the load is in
Conductance mode of operation. Conductance setpoint is stored in non-volatile memory; value is
restored upon power-up.
B.29 [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]? QUERY
COND?
Syntax:
Short Form:[SOUR:]COND[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]?
Returns
Description: Indicates conductance setpoint.
B.30 [SOURce:]CONDuctance:TRANsient COMMAND
COND:TRAN
Syntax:
Short Form: [SOUR:]COND:TRAN
Description: Configures a transient conductance pulse that is initiated by SYST:MODE:TRAN (see PAR.
B.100). See PAR. B.106 for details on using both Transient and Ramp commands at the same time.
Figure B-3 on page B-20 shows transient and ramp timing.